OTUS
Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 149
8-8-20-1
Removal Procedure
8-8-20-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Step
Corresponding graphic
1.
Lift the elevation arm to its highest position.
2.
Unscrew 2 screws to remove the Unity Elevation Arm Cover L/R
from the console
3.
Unscrew 2 screws to remove the Unity Elevation Arm Hide Cover
from the console.