Basic Service Manual
8 - 164
Section 8-9 - Replacing Front End Parts
8-9-3-2
Removal Procedure
8-9-3-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Steps
Corresponding Graphic
1.
Push the button of Unity Protector Assy’s connector
and disconnect the Unity Protector Assy’s Connector
from the console.
2.
Unscrew 6 screws to remove the Unity UCO-T Front
Cover Brkt.
3.
Unscrew 15 screws to remove the UCO-T from the
console.
Note:
If the UECG Assy was installed, remove the UECG
Assy as described in
.
If the U-DCWD Set was
installed, remove the U-DCWD Set as described in
Summary of Contents for LOGIQ Totus