LOGIQ T
OTUS
Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 155
8-8-24-2
Removal Procedure
Step
Corresponding graphic
1.
Unscrew 4 screws to remove the Unity Cable Cover
Brkt from the console
2. Unlock the Cable Clamp and disconnect the Fan
Cable from the Console
3.
Unscrew 2 screws to remove the Unity PWM Fan
Assy from the console