GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting
7 - 31
To configure automatic data transfers
:
1.) Navigate to
ADMIN > System Admin
.
2.) On
System Admin
, configure these settings:
-
Transfer Machine Logs to GE HealthCare
– Default is checked
-
Transfer Operator Information to GE HealthCare
-
Transfer Images (captured without patient information) to GE HealthCare
– Default is
checked.
3.) Click
Save
.
4.) Navigate to
Utility (second page) > Service > Utilities > Data Transfer
.
5.) On
Data Transfer
, select
Scheduler
, and then select the days to perform the data transfer.
6.) Click
Save Settings
.
7.) To manually perform a data transfer, click
Send All
.
For more information
:
•