GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
3 - 70
Section 3-7 - Setting Up Connectivity and Tips
3-7-6-6
Data Streaming
Use Data Streaming to stream live/recall/CINE ultrasound image data over the network connection to
enabled devices. The data stream will contain grayscale, color map, geometry, view settings (flip/rotate/
reverse), probe and system information VNav position information and CT/MR registration information
with ultrasound data. No patient information is transferred with the streamed data.
It is recommended to use a 1 Gbps network connection for Data Streaming. The required bandwidth
often lies in the 100-300 Mbps range. Usage of a 100 Mbps network leads to dropped frames and the
risk of latency buildup.
NOTE:
The Data Streaming option key needs to be installed to enable the Data Streaming page.
Table 3-32 To Configure Data Streaming
Step
Corresponding Graphic
1. Navigate to
Connectivity > Connection Manager > Special Devices
.
2. Select
Data Streaming
.
3. To allow data streaming, enable
Enable Streaming
.
4. To close data streaming when the patient/exam is changed, enable
Close Stream on Patient/Exam Change
.
5. In
PortNo
, enter the port number used for data streaming.
6. In Client Certificates, if needed, do one of the following:
- To remove selected client certificates, select
Revoke Selected
.
- To remove all expired client certificates, select
Revoke All Expired
.
7. Select
Done
.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......