GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
8 - 174
Section 8-10 - Replacing Back End Parts
8-10-4 Replacement of the COM EXPRESS MODULE, I5 CFL 16G
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
NOTE:
If you are replacing the SSD drive as FRU, the new SSD drive does not come loaded with
software, therefore a full software load is needed. See
Section 8-4 "Loading / Reloading /
Upgrading the Software" on page 8-8
.
8-10-4-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the Rear Cover Mid Assy-T as described in
8-8-10 "Replacement of the Rear Cover Mid
3.) Remove the Rear Cover Top Assy-U as described in
8-8-15 "Replacement of the Rear Cover Top
4.) Remove the T2 Side Cover L Assy as described in
8-8-7 "Replacement of the T2 Side Cover L
Table 8-93 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 15 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad