Section 8-4 - Loading / Reloading / Upgrading the Software
8 - 25
2.
With the trackball, point to 2 - (SYSTEM) and press the left
set key to select
2 - (System)
.
With the trackball, point to OK and press the left set key to
select
OK
.
NOTE
: Selecting
2-System
is a not DATA WIPE. ALL
EXISTING PATIENT DATA WILL BE PRESERVED.
This will take approximately 10-15 minutes.
NOTE
: If encryption is enabled, a warning message will
appear requesting to connect a USB keyboard to enter the
encryption recovery key. Select OK to continue and enter
the key using the alphanumeric keyboard.
3.
View the different screens that appear as the following
tasks are performed:
• Identifying Local Disk Properties
• Configuring Disk Partitions
• Populating Repository
• Populating System Drive. This may take some time...
• Populating Data Drives
4. When System Imaging is Complete displays, do the
following:
• Press OK to reboot the system.
• Remove the USB media BEFORE the system starts
rebooting.
Remove the USB media BEFORE the system starts
rebooting (right after the screens go black).
Table 8-9 Loading the base image and application software - System only(Continued)
Step
Corresponding Graphic
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......