Basic Service Manual
Section 8-7 - Replacing OPIO
8 - 111
8-7-19-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
NOTE:
Make sure of assembling T2 OPIO Key Rubber securely to OPIO PWA-T.
Figure 8-4 Example of good insertion