LOGIQ T
OTUS
Basic Service Manual
8 - 120
Section 8-7 - Replacing OPIO
8-7-24-2
Removal Procedure
Steps
Corresponding Graphic
1.
Unscrew 2 screw to remove the USB-C Cable.
Disconnect the USB-A and USB-C connector from
the console.
2.
Unscrew 2 screws to remove the User Port Module L
from the console.
3.
Unscrew 4 screws to remove the USB Port Fix Brkt
and remove the plastic clamp.
4.
Unscrew 2 screws to remove the User Port Module R
from the console.