Basic Service Manual
8 - 46
Section 8-5 - Replacing Cables
8-5-4-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
NOTE:
Fix OPIO HUB To Monitor Power Cable with Cable Tie as below figure.
-Align and fix the white double lined mark on the UBE To Monitor HDMI Cable and green label
attached on the OPIO HUB To Monitor Power Cable prior to second fixture using Cable Tie
(2EA).
-Fix the white double lined mark on the cable prior to second fixture using Cable Tie(2EA).
NOTE:
Align colored points