Section 8-7 - Replacing OPIO
8 - 89
8-7-2
Replacement of the OPIO Top Cover Set-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-7-2-1
Removal Procedure
8-7-2-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-38 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
No tools required
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Lift the OPIO Top Cover Left to remove from the console.
2.
Lift the OPIO Top Cover Right to remove from the console