Basic Service Manual
Section 8-11 - Replacing Options
8 - 195
8-11-10 Replacement of the T2 A/N Keyboard
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-11-10-1
Preparations
Remove the Unity Gel Warmer Cable Cover as described in
8-7-8 "Replacement of the Unity Gel
Warmer Cable Cover" on page 8-96
8-11-10-2
Removal Procedure
8-11-10-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-105 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
Phillips Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Disconnect 2 cables of T2 A/N Keyboard from the console.
2.
Unscrew 6 screws to remove the T2 A/N Keyboard from the
console.