LOGIQ Totus
Basic Service Manual
Chapter 10 Care & Maintenance
10 - 17
10-4-13
Air filter cleaning
Clean the system's air filters to ensure that a clogged filter does not cause the system to overheat and
reduce system performance and reliability. It is recommended the filters be cleaned every two weeks,
but the requirements will vary with environment.
1.) Pull Rear Cover Mid Assy-T to remove from the console.
Figure 10-3 Pull the Rear Cover Mid Assy-T to remove from the console
CAUTION
Be sure to lock the wheels before cleaning the air filters to avoid injury by any unexpected
movement of the system.
DO NOT operate the unit without the air filters in place.
Allow the air filters to dry thoroughly before re-installing them on the unit.