Basic Service Manual
8 - 58
Section 8-5 - Replacing Cables
8-5-9-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
NOTE:
Fix UBE To Monitor HDMI Cable with Cable Tie as below figure.
- Align the marks on UBE To OPIO Touch Panel HDMI Cable, GND Cable (Chassis To OPIO) and UBE
To Monitor HDMI Cable and fix the cables using Cable Tie, DACT-100A
-Align the marks on cables and fix the cables using Cable Clamp CPFDM14-A.
-Align and fix the white double lined mark on the UBE To Monitor HDMI Cable and green label attached
on the OPIO HUB To Monitor Power Cable prior to second fixture using Cable Tie (2EA).
-Fix the white double lined mark on the cable prior to second fixture using Cable Tie(2EA).