LOGIQ T
OTUS
Basic Service Manual
8 - 108
Section 8-7 - Replacing OPIO
8-7-17-2
Removal Procedure
8-7-17-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Step
Corresponding graphic
1.
Disconnect 1 cable from the T2 OPIO Lower Assy
2.
Unscrew 19 screws to remove the T2 OPIO Lower Assy.