LOGIQ Totus
Basic Service Manual
3 - 78
Section 3-7 - Setting Up Connectivity and Tips
3-7-9-5
Dataflow Notification
Use the Dataflow Notification page to set parameters for dataflow notifications.
3-7-9-6
Transfer Settings
Use the Transfer Settings page to set parameters for the data transfer.
Table 3-40 To Configure Dataflow Notification
Step
Corresponding Graphic
1. Navigate to
Connectivity > Connection Manager > Advanced Settings
.
2. Select
Dataflow Notification
.
3. To receive a warning when you press the Print key without an active patient, enable
Warn image store without patient
.
4. To post a warning message if you attempt to store images to a read-only Dataflow, enable
Warn image store to Read Only dataflow
.
5. To receive a warning when you register a patient to the “No Archive” data flow, enable
Warn register to No Archive
.
NOTE
: Select a different data flow for permanent storage of patient data.
6. To confirm action when ending an examination, enable
Request acknowledge of End Exam action
.
7. To post a warning if the video titles exist on the internal DVR flash memory, enable
Warn video titles exist in the internal storage
.
Table 3-41 To Configure Transfer Settings
Step
Corresponding Graphic
1. Navigate to
Connectivity > Connection Manager > Advanced Settings
.
2. Select
Transfer Settings
.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......