Section 8-7 - Replacing OPIO
8 - 95
8-7-7
Replacement of the T2 OPIO Base Rear Cover Assy
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-7-7-1
Preparations
1.) Remove the OPIO Rear Tray Set-T as described in
8-11-4 "Replacement of the OPIO Rear Tray
2.) Remove the Probe Holder Set-T as described in
8-7-3 "Replacement of the Probe Holder Set-T" on
8-7-7-2
Removal Procedure
8-7-7-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-43 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
Philips Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Unscrew 5 screws to remove the T2 OPIO Base
Rear Cover Assy from the console.