GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-11 - Replacing Options
8 - 221
8-11-24-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
5.
Unscrew 6 screws to remove the UBC-T Assy from
the console.
6.
Unscrew 8 screws to remove the Unity BBC Board Fix
Brkt.
Steps
Corresponding Graphic