GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
4 - 22
Section 4-3 - Functional Checks
4-3-10
System CFM and PWD functional checks
For information on the Ultrasound system’s different modes as well as expected results, the Basic User
Manual will familiarize you with image optimization for
B-Mode
,
M-Mode
,
Color Flow
, and
Doppler
.
For complete information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 5,
or the appropriate LOGIQ Totus Release Notes.
1-2-4-1 "LOGIQ Totus Options" on page 1-5
.
Preparations
You may use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner’s active probe connector.
2.) Turn ON the scanner.
4-3-11
Color Mode functional checks
Color Flow screens are B- or M-Mode screens with colors representing blood or tissue movement. Color
Flow may be selected both from B-Mode, or from M-Mode, or a combination of these screens.
For more information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 5, or the
appropriate LOGIQ Totus Release Notes.
4-3-12
Color B-Mode functional checks
For information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 5, or the
appropriate LOGIQ Totus Release Notes.
4-3-13
Color M-Mode functional checks
For information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 5, or the
appropriate LOGIQ Totus Release Notes.
•
Check Horizontal Sweep, PRF, Baseline, and Invert.
•
Check Variance, Color Maps, and Cineloop.
•
Check Frequency and Acoustic Power.
•
Adjust ROI.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......