Revision A
Dash 3000/4000 Patient Monitor
8-5
2000966-339
Field Replaceable Units and Upgrades:
Disassembly Guidelines
After Reassembly
After reassembling the monitor, ALWAYS complete the electrical safety
tests, checkout procedures, calibration tests, and regular maintenance
procedures identified and described in this manual.
Recommended Calibration, Electrical Safety Tests, and Checkout Procedures
Replacement or
Upgrade
Procedure
Calibration
Tests
Checkout Procedures
Electrical Safety Tests
Alarm light
option
Display
assembly
Display
assembly
components
Backlight
inverter PCB
None
“Monitor Power-up Tests” on page 4-
35
Complete Steps
of
the
“5 Lead ECG Test” on page 4-37
“Ground (Earth) Integrity” on page 4-26
“Ground (Earth) Wire Leakage Current Tests”
on page 4-28
“Enclosure Leakage Current Test” on page 4-
29
“Patient (Source) Leakage Current Test” on
page 4-30
“Patient (Sink) Leakage Current Test (Mains
Voltage on the Applied Part)” on page 4-31
Speaker
None
Complete ALL the Checkout Procedures
identified in Chapter 4,
.
Keypad
assembly
None
“Monitor Power-up Tests” on page 4-
35
GE DAS
assembly
Masimo DAS
assembly
“Respiration Tests” on page 4-40
“Temperature Tests” on page 4-41
“Cardiac Output Tests” on page 4-42
“Invasive Blood Pressure Tests” on
page 4-44
“Pulse Oximetry Tests for GE SPO2
Oximeter” on page 4-47
procedure if you installed a GE DAS
assembly.
“Pulse Oximetry Tests for Masimo SET
SPO2” on page 4-49
. Use this procdure
if you installed a Masimo DAS
assembly.
“Noninvasive Blood Pressure Tests” on
page 4-50
Refer to the CO2 chapter in the
monitor’s operator manual to checkout
the mainstream End-tidal CO2.
To check out the sidestream End-tidal
CO
2
, refer to the documentation that
came with the module.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......