5-12
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Troubleshooting:
Battery Alarms and Messages
Battery Alarms and Messages
Battery alarms occur when one of the following battery conditions occur:
Low battery
Empty battery
Battery failures
Charger failures
These Battery messages help identify the cause of a battery alarm.
Battery Messages appear in three different locations of the monitor:
ECG waveform area.
Battery Status Information window.
Battery Fuel Gauge icon.
Battery Messages Displayed in the ECG Waveform Area
The following battery messages are displayed in the ECG Waveform
Area.
Battery Messages Displayed in the ECG Waveform Area
Message
Battery Alarm
Cause
BATTERY LOW
System
WARNING
Critical Low Battery
Only 10 minutes per battery of run
time remaining (10 minutes if one battery, 20 minutes if two
batteries).
POWERING DOWN
System
WARNING
Empty Battery
There is no battery run time remaining.
CHECK BATT STATUS
System
MESSAGE
Battery Failure
A minor failure has occurred while using or
charging the battery.
BATTERY ERROR
System
WARNING
Battery Failure
A serious failure has occurred while using
or charging the battery.
CHECK BATT STATUS
NOTE
INTERNAL CHARGER
FAILED
,
CALL SERVICE
also appears in the Battery
Status information window.
System
MESSAGE
Charger Failure
Charger communications have failed.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......