Revision A
Dash 3000/4000 Patient Monitor
5-21
2000966-339
Troubleshooting:
Data Acquisition Tests
Generate Static BP Waveforms
1. Set the patient simulator BP output to 200 mmHg (240 mmHG
MarqII or MarqIII), static pressure.
2. Verify the BP channel is working correctly if systolic, diastolic, and
mean pressure values for BP are displaying parameter readings
between 194 and 206 mmHg.
3. Repeat steps 1 and 2 for each enabled BP channel.
BP Waveforms Do Not Appear Correctly On The Display
1. If the BP waveforms displayed on the monitor appear noisy or
distorted (example shown on the left), test the Patient simulator and
simulator test cables and on a working monitor to determine the
source of the problem.
2. If the static pressure test results were inaccurate, test the Patient
simulator and simulator test cables and on a working monitor to
determine the source of the problem.
3. If the patient simulator and associated test cables are determined to
be functioning correctly, replace the DAS FRU. See Chapter 8,
Replaceable Units and Upgrades”
BP Waveforms Do Not Appear On The Display At All
1. If the BP parameter labels, readings and associated waveforms do
not display on the monitor, verify the patient simulator and
associated test cables on a working monitor.
2. Inspect the BP side panel connectors on the monitor for bent or
broken pins. If any pins are bent or broken, replace the DAS
assembly.
3. Inspect the dual BP cable connectors on the monitor for bent or
broken pins. If any pins are bent or broken, replace the DAS
assembly.
520A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......