Revision A
Dash 3000/4000 Patient Monitor
6-7
2000966-339
Configuration:
Main Menu Selections
Set Graph Locations
Access Manual Graph Location option, starting from the Main Menu.
1. Select
MORE MENUS > MONITOR SETUP > GRAPH SETUP >
GRAPH LOCATION > MANUAL GRAPH LOCATION
.
2. Using the
Trim Knob
control, choose the manual graph location
from the list.
3. Select
ALARM GRAPH LOCATION
.
4. Using the
Trim Knob
control, choose the alarm graph location from
the list.
5. Select
PRINT WINDOW LOCATION
.
6. Using the
Trim Knob
control, choose the print window location from
the list.
7. Select
12 LEAD PRINT LOCATION
.
8. Using the
Trim Knob
control, choose the 12 lead print location from
the list.
Communication Confirmation
Confirm communication across the network.
1. Admit and generate a waveform at the monitor with a simulator.
2. Press
GRAPH GO/STOP
and observe graph output at chosen
locations.
Problems?
If the writer or printer does not graph:
Ensure the writer or printer is turned ON.
Check all cables for a good connection.
Check programmed alarms and manual graph locations at the
monitor.
If you do not have a waveform at the central station:
Ensure the central station software is compatible.
Check all cables for a good connection.
Check the programmed alarms and manual graph locations at the
monitor.
Ensure the care unit name is the same in the monitor and in the
central station.
Ensure the internet protocol (IP) addresses are programmed
correctly.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......