Revision A
Dash 3000/4000 Patient Monitor
8-19
2000966-339
Field Replaceable Units and Upgrades:
Display Assembly
Replacing the LCD Display
Complete the following steps for the monitor you are servicing.
Dash 3000 monitor
with display shield flex circuit
Dash 4000 monitor with Sharp display
assembly
1. Remove the handle assembly. Refer to page 8-7.
2. Remove the display assembly from the main unit. Refer to page 8-9.
3. Open the display assembly. Refer to page 8-16
4. Remove the LCD display. Follow the steps for the monitor you are servicing.
a. Peel back and remove the rubber
display isolator from around the
display.
b. Go to next step.
a. Disconnect the five cable connectors.
b. Remove the four screws anchoring
the mounting plate to the LCD display.
c.
Remove the mounting plate and set
aside.
504A
885A
885A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......