6-8
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Configuration:
Main Menu Selections
Admit Menu
The Admit Menu selection determines the function of the monitor. This
menu item is part of the Service Mode menu.
Before programming the Admit Menu, you must know if the monitor will
be used for standard adult, neonatal, or operating room monitoring, and
if the monitor will be moved from room to room. All combinations are
explained below.
Standard—configures the monitor to stay in one room for stationary
monitoring only. Monitors not connected to the Unity Network
(Ethernet connection) must use STANDARD configuration only.
Rover—configures the monitor to move from room to room for
stationary monitoring only.
Combo—configures the monitor to stay in one room for both
stationary and ambulatory (telemetry) monitoring. This monitor
displays all Tram module data combined with ECG data for
ambulatory patients.
NOTE
Users should be aware of a possible time discrepancy between the
waveforms from the Telemetry device and the waveforms hard-wired to
the Dash monitor. Users should not consider these waveforms to be
synchronous. If absolute synchronicity is desired, Combo mode should be
discontinued and the ECG waveforms should be acquired via the hard-
wired bedside monitor.
Rover Combo—configures the monitor to move from room to room for
both stationary and ambulatory (telemetry) monitoring.
1. Access Admit Menu option, starting from the Main Menu. Select
MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the
Trim Knob
control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407)
3. Select
MENU SETUP > ADMIT MENU.
4. Use the
Trim Knob
control to select the function of the monitor.
5. Press
Trim Knob
control to exit.
↓
↑
513A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......