Revision A
Dash 3000/4000 Patient Monitor
2-17
2000966-339
Equipment Overview:
Technical Specifications
GE Pulse Oximetry (SPO2)
Masimo SET SPO2 Module
Parameters monitored:
Arterial oxygen saturation (SpO2) and peripheral pulse
rate (PPR)
Measurement Range
SPO2:
PPR:
50 - 100%
30 - 300 beats per minute
Accuracy
SpO2:
PPR:
Actual accuracy depends on probe. Please reference
manufacturer’s specifications.
± 2% (70 - 100% SpO2) ±1 standard deviation
± 3% (50 - 69% SpO2) ±1 standard deviation
± 3 beats per minute
Alarms:
User-selectable upper and lower limits for SpO2 and
PPR.
Display Messages
LOW QUALITY, PROBE IS OFF THE PATIENT, PROBE OR
MODULE MALFUNCTION, PULSE SEARCH
Measurement Range
Saturation:
Pulse Rate:
Perfusion:
30 to 100% SpO
2
25 to 240 beats per minute
0.02 to 20%
Accuracy
Displayed saturation data
Saturation, no motion:
Saturation, motion:
Pulse Rate, no motion:
Pulse Rate, motion:
Functional saturation
SpO
2
over the range 70 to 100%, below 69% is unspecified
±2 digits for adults and pediatrics, ±3 digits for neonates
SpO
2
over the range 70 to 100%, below 69% is unspecified
±3 digits for adults, pediatrics, and neonates
25 to 240 bpm, ±3 bpm
25 to 240 bpm, ±5 bpm
Alarms:
User-selectable upper and lower limits for SpO2 and PPR.
Patents
5482036, 5490505, 5632272, 5685299, 5769785,
6002952, 6036642, 6067462, 5,758,644; 6206830;
6157850
.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......