6-12
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Configuration:
Boot Code Selections
Set Language
Select
Set Language
to change the language of the displayed text.
1. Activate the Boot Code.
a. Hold down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
2. Select the
SERVICE MENU
.
3. Select the
SET CONFIGURATION
menu option.
4. Select the
SET LANGUAGE
menu option.
5. Select the appropriate language.
6. Select
Exit
and wait for the monitor to display the message,
“Writing
settings to EEPROM...
”
7. Manually reboot the monitor to display the new language.
Enable/Disable AFIB Identification
Select
AFIB Identification
to enable or disable the atrial fibrillation
detection feature. This feature is part of the Cardiac software option.
NOTE
Country approval of the AFIB Identification software feature is
currently pending in Japan.
1. Activate the Boot Code.
a. Hold down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
2. Select the
SERVICE MENU
.
3. Select the
SET CONFIGURATION
menu option.
4. Select the
AFIB Identification
menu option.
5. Select
enabled
to activate this option, or
disabled
to deactivate this
option.
6. Select
Exit
and wait for the monitor to display the message,
“Writing
settings to EEPROM...
”
7. Manually reboot the monitor to implement this change.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......