Revision A
Dash 3000/4000 Patient Monitor
8-13
2000966-339
Field Replaceable Units and Upgrades:
Display Assembly
Replacing the Display Assembly
Dash 3000 Display Assembly
1. Remove the handle assembly. Refer to page 8-7.
2. Remove the display assembly from the main unit. Refer to page 8-9.
3. Replace the defective display assembly.
4. Reassemble the monitor in reverse order.
Dash 4000 with the Sharp Display Assembly
CAUTION
FLEX CABLE DAMAGE — DO NOT kink, pinch,
stretch, twist, or tightly fold a flex cable.
1. Remove the handle assembly. Refer to page 8-7.
2. Remove the display assembly from the main unit. Refer to page 8-9.
3. Remove the new display flex from the package and carefully install it
on the display assembly.
NOTE
You will need a thick book or some heavy object to support the
display assembly while you attach the display flex to the main
unit.
a. Insert the flex cable connector into the display connector as
shown in the figure below.
b. Using two of the flathead M3X6LG screws from the kit, secure
the flex cable to the display assembly.
873A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......