4-34
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Maintenance:
Checkout Procedures
Temperature Test
700/400 Series Dual Temperature Adapter (402015-004)
Temperature Simulator Cable for use with Marq-I only (6770031)
Cardiac Output Test
Cardiac Output Simulator II (900028-001) (no longer available for ordering)
Cardiac Output Cable Adapter (700092-001)
Masimo Sp0
2
Test
Masimo SpO
2
Test Kit (2021087-001). Includes Masimo Tester and SP02 Sensor
Adapter Cable.
NOTE
The Masimo Tester is intended for use only with Masimo
instruments or instruments containing Masimo SET oximetry
with standard Masimo PC Series patient cable connectors. See
the Masimo “Directions for Use” that came with the Tester for
more information
The PPR and SP02 values for the Masimo Tester must be
within these limits: Peripheral pulse rate: 61 bpm
±
1 bpm;
and SPO2 value: 81%
±
3%.
Clinical Dynamics’ SmartSAT
TM
Pulse Oximetry Analyzer (or equivalent). Contact
Clinical Dynamics for appropriate Nellcor or Masimo SET SpO
2
adapter cable.
However, the simulator comes standard with a Nellcor SpO
2
cable.
Fluke Biomedical’s Index 2
XL
SpO
2
Simulator (or equivalent). Uses either a Masimo
SET or Nellcor SpO
2
finger probe patient cable.
GEMS-IT Sp0
2
Test (Nellcor Probes)
GEMS-IT SpO
2
Simulator (408610-001 or equivalent).
NOTE
The GEMS-IT SPO
2
Simulator is only compatible with the
GEMS-IT SPO
2
parameter.
GEMS-IT SpO
2
Simulator Cable (700232-004 or equivalent)
Clinical Dynamics’ SmartSAT
TM
Pulse Oximetry Analyzer (or equivalent). Contact
Clinical Dynamics for appropriate Nellcor or Masimo SET SpO
2
adapter cable.
However, the simulator comes standard with a Nellcor SpO
2
cable.
Fluke Biomedical’s Index 2
XL
SpO
2
Simulator (or equivalent). Uses either a Masimo
SET or Nellcor SpO
2
finger probe patient cable.
NBP Test
NBP Cuff Coupling (400787-001)
NBP Hose Coupling (46100-002)
NBP Tee (4745-101)
NBP Tubing 2 feet (401582-001)
Manometer: digital (Sensym PDM200M), mercury, or equivalent
NBP Tube (414873-001)
NBP Cuff (9461-301)
Pipe: PVC
DEFIB Sync Test
Oscilloscope
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......