Revision A
Dash 3000/4000 Patient Monitor
8-15
2000966-339
Field Replaceable Units and Upgrades:
Display Assembly
4. Refer to the customer’s previous Dash 4000 display assembly to
determine if you need to enable or disable the alarm light.
a. To enable the alarm light option, verify the alarm light jumper is
connected to both J2 pins on the alarm light PCB.
b. To disable the alarm light option, remove the jumper from the
alarm light PCB, or connect the jumper to only one of the J2 pins
on the alarm light PCB.
5. Connect the display assembly to the main unit. Refer page 8-9 and
re-assemble the monitor in reverse order.
6. Using the labels from the kit, apply the Dash front panel and GE
“Dash 3000/4000 Assembly PN 2015999,
on page 9-33 for correct placement.
Replacing Display Assembly Components
Use the following procedures to replace the following display
components:
Backlight inverter PCB
Key pad assembly
LCD display
Dash 4000 alarm light
877A
clear lens (enable light)
opaque lens (disable light)
J2 alarm light jumper
878A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......