Revision A
Dash 3000/4000 Patient Monitor
7-5
2000966-339
Calibration:
NBP, Analog Output ECG, Analog Output BP, and End-tidal CO2 Software Calibration
The table below lists items for connecting the NBP tube between the
manometer and NBP cuff:
WARNING
When the NBP cuff is used in this procedure, it must be
tightly wrapped around a rigid cylinder or pipe. DO NOT
put the NBP cuff around a human arm during the
calibration procedures due to the potential for injury.
Calibration Procedure
Using the
Trim Knob
control, access the Service Mode menu starting
from the Main Menu.
1. Select
MORE MENUS-> MONITOR SETUP-> SERVICE MODE.
2. Enter password using the
Trim Knob
control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407).
3. Select
CALIBRATE-> CALIBRATE NBP-> CAL ZERO OFF->
START
.
4. The text on the menu item changes from Cal Zero Off to Cal Zero In
Progress.
When the process is complete, the menu item shows that it is
OFF
again.
3
Gain Calibration Test
NOTE
To proceed with the Gain Calibration Test, setup the monitor and
test equipment following the guidelines illustrated in the NBP
Checkout Procedure, Chapter 4, Maintenance.
1. Connect a cuff and manometer to the monitor.
2. Turn the digital manometer on and adjust the range switch to 1000
mmHg.
3.
Select
CAL GAIN OFF-> CAL GAIN OFF->
START
.
Description
Part Number
Qty
NBP cuff coupling
400787-001
1
NBP hose coupling
46100-002
1
NBP tee
4745-101
1
NBP tubing
401582-001
2
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......