Revision A
Dash 3000/4000 Patient Monitor
4-57
2000966-339
Maintenance:
Checkout Procedures
Network Test
1. Verify that the monitor is connected to the Unity-MC (Mission
Critical) network.
NOTE
When the monitor is connected to a docking station, verify the
docking station is connected to the Unity-MC network.
NOTE
When the monitor is connected to a docking station, the monitor’s
network port is disabled. Only the docking station’s network port
is enabled at this time.
2. Select
VIEW OTHER PATIENTS
.
3. Select
SELECT ANOTHER CARE UNIT
.
4. Verify that you can see at least one care unit.
5. Select a care unit.
6. Select
SELECT A BED TO VIEW
.
7. Select a bed.
8. Verify that the patient window appears on the monitor’s split-screen.
Remote Control Test (Option)
1. Verify the remote control is connected into an Autoport to Mport
adapter and is inserted into the
Aux
connector.
2. Select
MORE MENUS > MONITOR SETUP > REVISION AND ID.
3. Select
NEXT
from the popup menu to display the port connectors.
4. Verify the Remote Control label appears after the appropriate port
and the software version for the remote control is 1A.
5. Press each remote control key and verify a beep tone sounds at the
monitor.
RF LAN Test (option)
1. If the monitor has the Wireless LAN, disconnect the Ethernet cable
and verify Wireless LAN communication still exist between beds,
2. Reconnect the Ethernet Cable.
3. Return to the Main Menu.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......