Revision A
Dash 3000/4000 Patient Monitor
6-9
2000966-339
Configuration:
Boot Code Selections
Boot Code Selections
Set Defib Sync Voltage and Pulse Width
The monitor controls the analog out signal used to trigger a defibrillator.
Refer to the defibrillator manufacturer’s manual for the required pulse
amplitude and duration.
Use the Boot Code Service Menu to configure or change the Marker Out
signal of the
DEFIB SYNC
connector.
1. Activate the Boot Code.
a. Hold down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
2. Select
SERVICE MENU
.
3. Select
SET CONFIGURATION
menu option.
4. In the Configuration Menu, select:
1 Defib Sync Voltage:
and choose
5V
or
12V
amplitude.
2 Defib Sync Pulse Width:
and choose
10 ms
or
100 ms
for pulse
duration.
5. Select
Exit
and wait for the monitor to display the message,
“Writing
settings to EEPROM...”
6. Manually reboot the monitor to implement this change.
Set Line Frequency
Use the Boot Code Service Menu to configure or change the monitor line
frequency to 50 or 60 Hz. The default is 60 Hz.
1. Activate the Boot Code.
a. Hold down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
2. Select
SERVICE MENU
.
3. Select
SET CONFIGURATION
menu option.
4. Select
Line Frequency
then choose 50 Hz or 60 Hz line frequency.
5. Select
Exit
and wait for the monitor to display the message,
“Writing
settings to EEPROM...”
6. Manually reboot the monitor to implement this change.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......