9-44
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Assembly Drawings:
Port Connections
Masimo SuperStat Blood Pressures
Up to four invasive blood pressure channels are provided; each channel
uses a separate 11-pin, female connector. The pinout is as follows:
PIN
SIGNAL NAME
I/O
SIGNAL DESCRIPTION
DIAGRAM
1
BP1_+VREF or
BP2_+VREF
O
BP1 or BP2 transducer excitation voltage
2
BP SIG+
I
BP1 or BP2 transducer signal positive (+)
3
BP1_+VREF or
BP2_+VREF
O
BP1 or BP2 transducer excitation voltage
4
AGND
O
Isolated ground
5
INV or
INV
–
BP3 or BP4 transducer signal positive (+)
6
SHIELD
O
BP cable shield
7
ISO GND
O
Isolated ground
8
BP SIG1 or BP SIG2
I
BP1 or BP2 transducer signal negative (–)
9
ISO_ GND_MAIN
O
Isolated ground
10
BP1_ID or BP2_ID
I
BP1 or BP2 probe identification signal
11
INV_BP3SIG or
INV_BP4SIG
–
BP3 or BP4 signal negative (–)
868A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......