Revision A
Dash 3000/4000 Patient Monitor
4-49
2000966-339
Maintenance:
Checkout Procedures
Pulse Oximetry Tests for Masimo SET SPO2
1. Connect the Masimo-style SPO2 simulator cable between the SPO2
connector of the monitor and the SPO2 simulator.
NOTE
Do not connect the simulator to an AC power supply for these tests.
Operate the simulator on battery power.
2. Set the simulator as follows:
a. Turn the monitor on and the simulator power
ON
.
b. Set the simulator to be compatible with Masimo SET SPO2. For
specific set up options, see the documentation provided with the
patient simulator.
c. Set the SPO2 % to
81
.
d. Set the PRR to
61
beats/minute.
3. Verify the following are displayed at the monitor: (It might be
necessary to turn the SPO2 parameter on.)
A waveform with an SpO
2
label.
An SPO2 % reading between
78 - 84
%.
A PRR reading between
58
and
64
beats per minute.
4. Create an alarm condition by setting the SPO2 low limit to
90
%.
5. Make sure the SPO2 % value on the monitor flashes, and it sounds
an alarm.
6. Return the low limit to the original value.
7. Disconnect the simulator cable from the module.
824C
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......