9-14
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Assembly Drawings:
Theory Of Operation
Microprocessor Supervisory Circuit, Microcontroller Internal Watchdog Timer
The microprocessor supervisory circuit provides reliable operation of the
main processor board. This circuit monitors the +3.3 V power supply and
asserts a 140 mS active low reset pulse when the power supply voltage is
below +3.0 V during power-up and power-down conditions. A 3 V, 0.5 A
hour lithium battery is used to preserve the contents of two SRAM
devices and a real-time clock (RTC) when VCC is below the reset
threshold.
System Control Logic
The system ASIC contains all of the system control logic for the
processor/power management PCB. Such functions include address
decoding, peripheral read and write control strobes, smart battery
control logic, display control, multiple I/O ports, and front panel key
switch debouncing.
Memory
Eight Mb of non-volatile memory are provided to support the boot code
and expansion memory such as high resolution graphic trends data
storage. The 512 Kb boot block is write protected.
Four Mb of non-volatile memory are provided to support the main
software application code.
Four Mbytes of volatile memory are provided by the synchronous
graphics RAM. This memory is used for stack, variable storage,
dynamically allocated memory and video data storage.
One Mb of battery-backed SRAM supports storage of 24 hours of 1-
minute resolution patient trends, an error log containing 50 input errors,
and 50 output errors and storage for the CPM buffers.
The monitor configuration data such as Internet and Ethernet addresses,
unit name and bed number are maintained in the 2 Kb EEPROM. The
serial device resides on the I
2
C bus.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......