Revision A
Dash 3000/4000 Patient Monitor
4-33
2000966-339
Maintenance:
Checkout Procedures
Checkout Procedures
Manufacturer Recommendations
These checkout procedures provide service personnel with a method to
verify operational and functional performance of the monitor. Failure to
attain any of the listed results indicates a potential malfunction of the
monitor.
The checkout procedures are based on the assumption that the tested
monitor has known good cables and test equipment. It also requires that
the user be familiar with the operation of all test equipment required for
the checkout procedures. For more information concerning the operation
of these components, refer to the respective operator manual(s).
Frequency
GE recommends that you perform the checkout procedures:
Upon receipt of the device.
Every 12 months thereafter.
Each time the main enclosure is disassembled or a circuit board is
removed, tested, or replaced.
Test Equipment
The following table lists GE recommended test equipment, adaptors, and
cables you need to successfully complete the checkout procedures. The
checkout procedures are written for the test equipment in the following
table. If you use test equipment other than those GE recommends, you
may need to slightly modify some test steps.
Patient Simulators and Cables
ECG Test
Marq-I ECG Simulator (no longer available for ordering)
MarqII-KIT (includes CO injectate box. No longer available for ordering)
MarqIII-KIT (includes CO injectate box)
NOTE
You can use a MarqI, MarqII, or MarqIII patient simulator or an
equivalent 12SL ECG patient simulator.
ECG Patient Cable
ECG Leadwire Set
Blood Pressure Test
Blood Pressure Simulator Adapter Cable (700095-001)
Dual BP cable
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......