Revision A
Dash 3000/4000 Patient Monitor
2-7
2000966-339
Equipment Overview:
Components
Function Keys
Fixed Keys
Four fixed function keys are provided for
GRAPH GO/STOP
,
NBP GO/
STOP
,
ZERO ALL
, and
SILENCE ALARM/ADMIT
.
Indicators
While the monitor powers up or changes between normal mode and
standby mode, all four front panel indicators illuminate.
AC Power Indicator
The indicator illuminates green when AC mains power is applied to the
monitor (including when the monitor is in the standby mode). The
indicator does not illuminate when the monitor is not powered by AC
mains power.
Battery Power Indicator
The indicator illuminates yellow when the monitor is operating on
battery power. The indicator does not illuminate when the monitor is not
battery powered.
Charging Status Indicators
The following table explains what the Charging Status indicators mean.
“Battery In Use” Indicators
The “Battery In Use” indicator (inside the battery door) illuminates
green when the monitor is receiving power solely from the respective
battery. The indicators do not illuminate when the monitor is not battery
powered.
Neither indicator illuminates when the monitor is operating from both
batteries simultaneously (i.e., in a very low battery charge condition
when both batteries are joined together in order to sustain operation of
the monitor).
LED Color
Explanation
Yellow
Two battery icons, labeled
Charging Status A
and
B
, illuminate
yellow when the respective battery is being charged. If both batteries
are present and require charging, then both icons illuminate yellow
even though they charge sequentially.
Green
The icon illuminates green when the respective battery is fully
charged.
No Light
The icon does not illuminate under the following conditions:
The respective battery is not installed.
The monitor is operating on battery power.
A failure condition has been detected for the respective battery.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......