Revision A
Dash 3000/4000 Patient Monitor
6-15
2000966-339
Configuration:
Advanced User Procedures
Advanced User Procedures
The following procedures are for advanced users only. These procedures
should rarely be used, and only experienced technicians should proceed.
Procedures
The following procedures are discussed later in this chapter.
Set Time and Date
Transfer Monitor Defaults
Change Ethernet Address
Set Internet Address
Reviewing Error Logs
Transferring Error Logs
Reviewing Event Logs
Set Time and Date
Change the time only when the system is switched to or from daylight
savings time.
NOTE
When a monitor is first connected to the optional Unity Network, the
time and date is automatically updated from the network time.
WARNING
Loss of patient data history. Changing the time or date
settings may result in the loss of patient data history. If
one monitor’s time or date is changed, all monitors on the
network “listen” and follow suit within 3-5 seconds.
Changing the time base of one monitor may cause some
loss of patient data history for all the monitors on the
network.
The following procedure explains how to use the TIME AND DATE
option in the monitor SERVICE MODE menu.
1. Access the Time and Date menu starting from the Main Menu. Select
MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the
Trim Knob
control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407)
3. Select
SET TIME
and use the
Trim Knob
control to change the
time. The time displays as a 24-hour military clock.
4. Select
SET DATE
and use the
Trim Knob
control to change the
date.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......