4-20
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Maintenance:
Battery Maintenance
Waking Up the Battery With a Monitor
1. Connect the monitor to an AC power source.
2. Activate the Boot Loader program.
a. Hold down
NBP GO/STOP
and
ZERO ALL
(
FUNCTION
).
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
(
FUNCTION
)
until the Boot Code information appears on the display.
3. Select
Wake Up Battery
from the Service Menu.
4. Insert the “sleeping” battery into slot
A
of the monitor.
NOTE
In some languages, slot
A
is identified as slot
1
and slot
B
is
identified as slot
2
.
5. Select
Wake Up Lithium Ion Battery in SLOT A
. Then, the monitor
applies a “wake-up” shot of current for up to 210 seconds.
6. Watch for the message,
Battery awake, run main code to charge
. The
battery should now be “awake.”
If the
“Battery awake, run main code to charge
” message appears
but the battery is NOT awake, the battery may be in a deep
discharge state. You should repeat step 5 at least four more times
to give the battery additional “wake-up” shots of current.
7. Quickly reboot (or power cycle) the monitor to begin charging the
batteries. The monitor’s illuminated yellow-colored
CHARGING
STATUS
LED indicates that the battery is being charged.
NOTE
The monitor will not charge the battery while it is running the
Boot Loader program.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......