Assembly Drawings:
Port Connections
Peripheral Expansion Interface Connector
An 18-pin peripheral expansion interface is provided to support future
use. An active low peripheral present signal may be polled by software to
identify when a peripheral is attached to the monitor. Asynchronous,
Aux, and a switched Ethernet serial communication channels are
supported as well as sw9-18V and +5V power.
PIN
SIGNAL NAME
I/O
SIGNAL DESCRIPTION
1
RETURN
–
Common power return
2
+9-18V
O
+9-18V power
3
PER_MARKER_OUT
O
Marker out
4
PER_ENET_PRESENT*
I
Signal to request switched Ethernet to peripheral
5
PER_ENET_TXD–
O
Transmit data + to peripheral
6
PER_ENET_RXD–
O
Transmit data – to peripheral
7
PER_E
I
Receive data + from peripheral
8
PER_E
I
Receive data – from peripheral
9
PER_AUTOPORT_ID_RXD
I
AutoPort ID RXD
10
PER_PRESENT*
I
Asserted by the installed peripheral device once powered
11
PER_ASYNC_RXD
I
Serial asynchronous data input from peripheral
12
PER_ASYNC_TXD
O
Serial asynchronous data output to peripheral
13
TC_PACER_BLANK*
O
Pacer blanking pulse from defibrillator
14
PER_AUTOPORT_ID_TXD
O
AutoPort ID TXD
15
PER_AUTOPORT_RXD
O
AutoPort RXD
16
PER_AUTOPORT_TXD
I
AutoPort TXD
17
+5V
O
+5V power
18
RETURN
–
Common power return
573B
Bottom View of the Monitor
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......