8-44
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Field Replaceable Units and Upgrades:
Optional DDW Writer Replacement/Upgrade
Optional DDW Writer Replacement/Upgrade
Replacement
There are two compatible writers. The instructions for assembly/
disassembly are the same for each writer.
NOTE
You do not need to disassemble the monitor to replace or upgrade the
optional printer.
1. Remove the writer by unscrewing the two captive screws inside the
writer.
CAUTION
Make sure you approach the screw from
below
the top bar
on the paper roll holder so that you do not damage the
unit.
2. Remove the writer from the unit.
3. Insert the new writer into the rear housing to engage with the blind
mate connector.
4. Secure the writer with the two captive screws inside the writer.
5. Perform the Graph Test in chapter four, Maintenance.
Upgrade
If upgrading the monitor to add a writer, pry off the blank cover with a
flat blade screw driver. The cover may crack or break when removing.
Once the cover is removed it can be discarded. Proceed with steps 3 and 4
above.
2-Inch Writer Assembly FRU
546A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......