Revision A
Dash 3000/4000 Patient Monitor
2-13
2000966-339
Equipment Overview:
Technical Specifications
Alarms
ECG
Categories:
Patient Status and System Status
Priorities:
4 levels — Crisis, Warning, Advisory, and Message
Notification:
Audible and visual
Setting:
Default and individual
Silencing:
1 minute, current alarm only
Pause:
5 minutes (adult); 3 minutes (neonatal); 5, 15 minutes,
permanent (OR mode)
Volume:
Default 70 dB measured at 1 meter
5 Leadwire cable:
I, II, III, V, aVR, aVL, and aVF
10 Leadwire cable (12SL option): V2, V3, V4, V5 and V6
Leads analyzed simultaneously:
I, II, III, and V (multi-lead mode)
Lead fail:
Identifies failed lead
Alarms:
User-selectable upper and lower heart rate limits
Input specifications:
Voltage range:
Signal width:
Heart rate range:
Accuracy:
Input impedance:
Common mode:
Differential:
Common mode rejection:
±0.5 mV to ±5 mV
40 ms to 120 ms (Q to S)
30 to 300 BPM
±1% or ±1 BPM, whichever is greater
>10 M
Ω
at 50/60 Hz
>2.5 M
Ω
from dc to 60 Hz
90dB minimum at 50 Hz or 60 Hz
Output specifications:
Heart rate averaging:
Frequency response:
Display:
Diagnostic Mode:
Monitoring Mode:
Moderate Mode:
Maximum Mode:
Impulse Response:
8 beats with a 2-second update rate of display
Response of non-permanent displays is limited by
resolution to 40 Hz (-3dB) @ 25 mm/s. Specified upper
frequency limits may vary by ± 2 Hz.
0.67(+0.4 dB) to 100 Hz (-3 dB)
For compliance with China National Standard: 1.0 Hz
(+0.4 dB) to 75 Hz (-3 dB)
0.67 (+0.4 dB) to 40 Hz (-3 dB)
0.67 (+0.4 dB) to 25 Hz (-3 dB)
5.0 Hz (-3 dB) to 25 Hz (-3 dB)
For an impulse of 3 mV applied for 100 ms:
Displacement following impulse: < 0.1 mV
Slope following impulse: < 0.3 mV/s
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......