9-20
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Assembly Drawings:
Theory Of Operation
Handle Subassembly
The handle serves multiple purposes in the monitor. The modular design
enables the user to add adapters for specialized applications.
The Dash 3000 handle houses the optional alarm light. This light is
visible for 360° surrounding the monitor. The light is intended for
applications when the audible notification is not useful or effective, such
as noisy environments (e.g., emergency vehicles) or quiet environments
(e.g., neonatal care areas). The alarm light indicates two levels of visual
alarms:
Crisis alarms (red LED) and
Warning alarms (amber LED).
The handle also houses the two RF LAN antennas required for the
optional RF LAN subsystem.
Interfaces
Ethernet
The Ethernet RJ-45 connector provides a hardware connection to the
optional Unity Network. The monitor has a built-in transceiver for
twisted-pair wire. Basic insulation (1500 VAC) isolates the monitor from
networked devices.
AUX
The AUX RJ-45 connector provides an asynchronous communication
connection to devices within the bedside care area. Basic insulation (1500
VAC) isolates the monitor from other devices.
Defib Sync
The Defib Sync connector provides signals needed to perform
synchronized cardioversion with a defibrillator. The Marker Out signal is
a pulse with selectable amplitude and width that coincides with the
patient’s ECG R-wave. The Marker In signal is returned to the monitor
by the defibrillator. The Marker In signal causes the monitor to insert a
defib marker in the displayed ECG waveform.
This connector also provides two analog signals: ECG and invasive
pressure. The monitor provides the top displayed ECG signal with
reconstructed pace pulses. You can use this signal to trigger a
defibrillator or intra-aortic balloon pump. BP1 produces the pressure
signal and is intended for triggering an intra-aortic balloon pump.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......