Revision A
Dash 3000/4000 Patient Monitor
9-15
2000966-339
Assembly Drawings:
Theory Of Operation
Real-Time Clock
The real-time clock incorporates an on-board quartz crystal. This feature
simplifies the design and eliminates adjustments. The time of day is
maintained to an accuracy of 15 seconds in 24 hours to support 24 hour
trend data. The RTC device is one of the devices on the I
2
C bus.
Audio Subsystem
Audio tones are generated using a tone generator, an audio amplifier,
and an 8
Ω
, 2.5-in. speaker. The tone generator has built in D/A
converters and a mixer to generate the dual frequency tones.
Frequencies ranging from 150 Hz to 2800 Hz are produced.
Video Subsystem
The microprocessor contains a video controller that supports real-time
and non-real-time waveform drawing, menu drawing, and parameter
display.
Display brightness is controlled by a 50 k
Ω
digital potentiometer. The
potentiometer is interfaced to the brightness control input of an inverter.
Analog Outputs
Two analog output channels support ECG and BP. The pace pulse is
generated on the main processor board and inserted into the analog out
ECG signal. The digital-to-analog conversion for both ECG and blood
pressure output signals are performed on the main processing board via
a two channel 12-bit serial DAC. Cal data is stored on the processor/
power management PCB.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......