Calibration:
NBP, Analog Output ECG, Analog Output BP, and End-tidal CO2 Software Calibration
NBP, Analog Output ECG, Analog Output BP, and
End-tidal CO2 Software Calibration
Noninvasive blood pressure (NBP), Analog Output ECG, Invasive
Pressures (Analog Output BP), and End-tidal CO2 (optional) require
software calibration.
Perform the NBP and End-tidal CO2 software calibration upon receipt of
the equipment, every 12 months thereafter, and each time the unit is
opened for service. This ensures the pneumatic circuit plumbing has not
developed any air leaks as a result of disassembly. Perform the analog
output ECG and analog output BP software calibration only if these did
NOT pass the analog output checkout procedure.
NBP Calibration
In General
The overall accuracy of noninvasive blood pressure (NBP) readings by
the monitor depend on the following:
the zero pressure reading, and
the voltage span of the NBP sensor in the monitor.
This procedure provides a method of verifying these items are accurate
and also checks the NBP pneumatic circuit plumbing for leaks.
The manufacturer recommends performing this procedure upon initially
receiving the monitor, before it is used on a patient, and once each year
thereafter. Also, perform the procedure each time the monitor is opened
for service or repair, simply to verify the NBP pneumatic circuit
plumbing did not develop inadvertent air leaks.
Required Test Equipment
The following items are required to successfully complete the NBP
calibration procedure.
Refer to
“Noninvasive Blood Pressure Tests”
on page 4-50 for setup
guidelines.
Voltmeter,
Unterminated defib sync cable,
Manometer (Sensym PDM200M or mercury manometer),
NBP tube, pn 9461-203,
NBP cuff, pn 9461-301 (any size works), and
Something to wrap the NBP cuff around (PVC pipe or coffee can).
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......