Field Replaceable Units and Upgrades:
Display Assembly
Opening the Display Assembly
Complete the following steps for the monitor you are servicing.
Dash 3000 monitor
with display shield flex circuit
Dash 4000 monitor with Sharp display
assembly
Separate the monitor bezel from the LCD
display.
1. Remove the four screws holding the
display shield to the LCD display and the
two screws holding the display shield to
the monitor bezel.
2. Remove the display assembly from the
monitor bezel.
3. Pull the LCD display isolator out of the
way. Disconnect the LCD display cable.
501A
4. Disconnect the cable from the backlight
inverter PCB.
502A
Separate the transition bezel from the front
bezel.
1. Remove the two screws from the display
flex cable. Pull up on the latch taps to
remove the flex cable from the display.
2. Remove the six screws from the back of
the transition bezel and the four screws
from the back of the metal display plate.
3. Separate the transition bezel from the
front bezel.
4. Set aside the transition bezel and the
metal display plate.
two smaller screws
four larger screws
500A
LCD display cable
cable
879A
2 screws
six screws
four screws
880A
884A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......