Revision A
Dash 3000/4000 Patient Monitor
4-37
2000966-339
Maintenance:
Checkout Procedures
ECG Tests
5 Lead ECG Test
Perform this test if the monitor acquires 5 lead ECG data.
1. Set up the patient simulator as follows.
Heart rate – 80 bpm.
Heart rate amplitude – 1.0 mV.
5-lead ECG patient cable properly attached.
2x gain for a MarqI, MarqII, or MarqIII simulator.
2. Attach the ECG patient cable and ECG leadwire set to the
ECG/
RESP
connector on the monitor and the leadwire connectors on the
top of the patient simulator.
3. Admit the patient simulator to the monitor.
4. Observe the following:
ECG lead II is displayed and is noise-free,
Heart rate of 80 ±1 bpm is displayed,
With QRS tones enabled, an audible tone sounds with each R-
Wave (QRS complex).
5. Verify all six ECG leads are available to view and are noise-free.
6. Select
DETECT PACE
and set to
PACE 2
.
7. Select a pacemaker pulse on the simulator.
015D
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......