4-42
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Maintenance:
Checkout Procedures
Cardiac Output Tests
Cardiac output testing is different depending on the simulator you use.
The Cardiac Output Simulator II is no longer available, but if you
already have that simulator, use the first section to run this test.
Otherwise, go to “Test with Marq II or MarqIII Simulator.”
Test with Cardiac Output Simulator II
1. Connect the cardiac output (CO) cable adaptor to the
TEMP/CO
connector of the monitor.
2. Connect a simulator cable between the CO cable adaptor and the CO
simulator.
3. Set the CO simulator to output blood temperature (BT) readings, as
found in the following table:
4. Verify a CO parameter window appears on the monitor display with
correct BT readings as shown in the table above.
Simulator BT Setting
Monitor BT Reading Range
30.3°C
30.1 – 30.5
35.1°C
34.9 – 35.3
36.0°C
35.8 – 36.2
37.0°C
36.8 – 37.2
41.7°C
41.5 – 41.9
018A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......