Revision B
120 Series Maternal/Fetal Monitor
10-15
2015590-001
Strip Chart Recorder Servicing: Field Serviceable Assemblies
Adjusting the Paper-Loading Sensor
1.
Load paper in the recorder. (Refer to
page 5-2
for proper loading instructions—
paper should feed from the
top
of the pack.) Ensure that there are
no
black
squares showing to indicate a paper-low condition. You must load at least nine
sheets of paper.
2.
If you haven’t already, re-connect the power cord and turn on the monitor.
3.
Using a digital voltmeter, connect the positive lead to J9 (pin 6) on the Recorder
Board; connect the negative lead to J9 (pin 2).
4.
Adjust R41 for a reading of +190 mVdc ± 5 mVdc on the digital voltmeter.
5.
Although a factory-only fixture is required for this step, you may wish to try it
by estimating the required distance. Load paper
backwards
(paper feeding from
bottom
of pack) with the white surface 0.380 in away from the paper tray. This
creates a paper-loading error condition. Verify that the voltage at J9 (pin 6) on
the Recorder Board is at least +2.70 Vdc.
6.
Replace the monitor top cover and secure with all nine screws.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......